Chlorine dioxide analyser M 1056 C

Measuring range: 0 - 1 mg/l

  • Platinum/ Platinum measurement electrodes, reference electrode
  • Online chlorine dioxide measurement
  • Stable measurement— open flow measurement cell<&l

    • 3 measuring electrodes platinum/platinum + reference electrode
    • Continuous measuring: chlorine doixide
    • Measuring range for concentrations up to 500 ug/l 
    • Stable measurement— open flow measurement cell with a constant flow
    • Continuous cleaning of measuring electrodes by stepper motor
    • Insight in measurement cell parameters for rapid detection of interfering substances
    • Modular electronics for simple service
    • Alarm for measuring cell or regulation errors
    • CANbus communication to easily connect several units to the control unit
    • Clear graphical display for all measured parameters

    GENERAL

    Chlorine dioxide analyser series M 1056 C is designed for measurement and control of drinking water, waste water treatment and swimming-pool applications.
    Two basic measured parameters are chlorine doixide and temperature.
    The unit enables the construction of additional expansion modules, for measuring the pH value and the redox potential (for drinking and waste water).
    Amperometric measurement method using two electrodes and an additional third reference electrode provides a constant and continuous measurements.
    The unit consists of the following subassemblies:
    - a mechanical filter sample
    - an indicator of the flow of the sample
    - self-cleaning measuring cell
    - measurement - regulation electronics
    - cover

    The user interface for parameterization and calibration is performed via graphical LCD display. Control, regulation and control of disinfection processes are enabled by analog and digital outputs and by CANbus digital communication. The control module supports the connection of actuators - dosing pump with analog and digital inputs as well as solenoid and motorized valves - for two-point and continuous dosing. Software design of the control module contains a number of protective mechanisms against overdosing.

    Contact us for more info

    General data:

    Ambient temperature: -10...+50ºC

    Relative humidity: 10...95 % non-condensating

    Control unit protection: IP 65

    Dimensions: 670 x 322 x 113

    Weight: 6,5 kg

    Flow: minimum 0,5 l/min

    Measuring range (M.R.): 0 - 1 mg/l

    Measuring principle: (amperometric): Platinum-platinum electrodes and third reference electrode 

    Measured value resolution: 0,001 mg/l

    Deviation of indication, measured value: maximum 2% of M.R.

    pH range: 6...10

    Water conductivity: > 50µS

    Automatic temperature compensation range: 0...+50ºC

     

    Electrical data:

    Supply voltage: 24 VAC/VDC ±20%

    Power: 5 W Maximum

    Isolation voltage: 500V (to analogue part)

     

    Current output connector:

    Modules: 2

    Current range: 4...20 mA

    Load: maximum 800 E at 24VDC

    Output range: adjustable to (MR)

    Output Mode options: - Free chlorine signal output

    - Temperature

    - Regulator output

    Regulator connector:

    Regulator 3 point or PI regulator

    Outputs: 3 x 24 VAC / 250 mA

    Inputs: 3 x 24 VAC

    Mode options:
    - Motor Control
    - Pump Control